beautypg.com

5putting into operation – BUCHI Inert Loop B-295 User Manual

Page 27

background image

5 Putting into operation

27

B-290 Operation Manual, Version I

5

Putting into operation

This chapter describes how the instrument is installed and gives instructions on initial startup.

NOTE
Inspect the instrument for damages during unpacking. If necessary, prepare a status report immedi-
ately to inform the postal company, railway company or transportation company.

Keep the original packaging for future transportation.

5 .1

Installation site

Put the instrument on a stable, horizontal surface. Consider the maximum product dimensions and
weight and the operation height of 1.30 meters for spray chilling applications. Make sure that you can
comfortably reach the instrument.
Installation on a mobile trolley is also possible. Obtain the environmental conditions as described in
section 3.2, technical data.

Put the instrument on a stable, horizontal base.

!

Warning

Death or serious injuries by use in explosive environments.
• Do not operate the instrument in explosive environments
• Do not operate the instrument with explosive gas mixtures without inertization
• Before operation, check all gas connections for correct installation
• Directly withdraw released gases and gaseous substances by sufficient ventilation

!

Caution

Risk of minor or moderate injury by heavy weight of the instrument.
• Consult a second person to transport the instrument
• Do not tilt or drop the instrument
• Place the instrument on a stable, even and vibration-free surface
• Keep limbs out of crushing zone
• Lift the device at the right and left bottom reaching areas for the hands

5 .2

Electrical connections

The Mini Spray Dryer B-290 is aimed to be installed mobile and is and is shipped with a flexible power
cord. After the installation, an electrical safety test is recommended to verify a safe system condition
such as sufficient grounding.

NOTICE

Risk of instrument damage by wrong mains supply.
• External mains supply must meet the voltage given on the type plate
• Check for sufficient grounding

This manual is related to the following products: