6 .2 conversion tables for the parameters – BUCHI Inert Loop B-295 User Manual
Page 49
![background image](/manuals/656064/49/background.png)
6 Operation
49
B-290 Operation Manual, Version I
6 .2
Conversion tables for the parameters
6 .2 .1
Flow meter spraying air (rotameter)
The rotameter is an indicator for the spray gas flow. The table gives a correlation between indicated
height and volume throughput. The nozzle has a certain pressure drop which increases with higher
gas flow. As the gas volume strongly corresponds to the actual pressure, the table also contains a row
for the effective volume flow, determined in a spray process with air.
Height (mm)
Liters/hour
Pressure drop (bar)
Actual volume flow (at standard
temperature and pressure)
in liters/hour
5
84
10
138
15
192
20
246
0.15
283
25
301
0.18
355
30
357
0.23
439
35
414
0.3
538
40
473
0.41
667
45
536
0.55
831
50
601
0.75
1052
55
670
1.05
1374
60
742
1.35
1744
65
819
1.8
2293
6 .2 .2
Peristaltic pump
The peristaltic pump can be adjusted to dif-
ferent tubes according to the inner and outer
diameters. With different tube diameters, the
absolute flow is varying. The diagram shows
the correlation for the standard silicone tube
(2 mm ID, 4 mm OD) for water.
Fig. 6.3: Pump settings versus throughput