Warning – BUCHI Inert Loop B-295 User Manual
Page 40
5 Putting into operation
40
B-290 Operation Manual, Version I
!
Warning
Death or serious poisoning by contact or incorporation of harmful substances at use.
• Before operation, check the instrument for correct assembling
• Before operation, inspect sealings and tubes for good condition
• Exchange worn out or defective parts immediately
• Exchange clogged filters immediately
• Only operate the instrument in ventilated environments
• Directly withdraw released gases and gaseous substances by sufficient ventilatio.
• Perform a dry-run without sample material and check for gas leakages
Fig. 5.18: Inert Loop B-295
The Inert Loop B-295 has wheels and can be placed
next to the Mini Spray Dryer B-290. Moreover it can be
used as a trolley if the instrument has been purchased in
2011 or later.
Connect the system to the mains supply with the power
cord.
Connect the two instruments with the communica-
tion cable to ensure their safe operation. The tubing is
described below.
Fig. 5.19: Mini Spray Dryer B-290 Advanced with Inert Loop B-295 as trolley