2 Safety

11

B-290 Operation Manual, Version I

!

DANGER

Death or serious poisoning by gases or particles due to O

2

- sensonr or filter malfunction

• Exchange defective O

2

-sensor immediately

• Exchange O

2

-sensor regularly within the specified maintenance intervals

• Exchange clogged filters immediately
• Exchange filters regularly within the specified maintenance intervals
• Dispose of filter immediately

!

DANGER

Death or serious poisoning by inhalation or incorporation of dried particles during spray process.
• Wear safety goggles
• Wear safety gloves
• Wear a suitable protective mask
• Wear a laboratory coat
• Check for proper sealing before use
• Do not inhale dried particles
• Stop drying gas flow before opening the drying circuit