5 Putting into operation

43

B-290 Operation Manual, Version I

Used as inlet air conditioning system
Installation steps:
1. Place the Dehumidifier B-296 next to the Mini Spray Dryer B-290.
2. Connect the tubing a from the outlet (labelled on the enclosure) of the Dehumidifier B-296 to the

gas inlet of the Mini Spray Dryer B-290 heater.

3. Use hose clamps to fix the tubing.
4. Connect the power cord to the mains. The device can now be switched on.

a Dehumidifier B-296 outlet tube

Fig. 5.23: Tubing between the Mini Spray Dryer B-290 and the Dehumidifier B-296

Used in the closed mode
When mixtures of organic solvents and water (with less than 50 % solvents) are being processed in
‘closed loop’ configuration, the use of the Dehumidifier B-296 is highly recommended. The cooling
temperature is fixed to 0 °C.

a Inlet tubing B-296
b Outlet tubing B-296

Fig. 5.24: Tubing between the preheat exchanger and the Mini Spray Dryer B-290

Installation steps:
1. Place the Dehumidifier B-296 next to the Mini Spray Dryer B-290.
2. Connect the tubing b from the outlet (labelled on the enclosure) of the Dehumidifier B-296 to the

gas inlet of the Mini Spray Dryer B-290.

3. Connect the tubing a from the Mini Spray Dryer B-290 outlet to the inlet of the Dehumidifier

B-296 (labelled on the enclosure).

4. Use hose clamps to fix all Polypress tubings.
5. Connect the power cord to the mains. The device can now be switched on.