5 Putting into operation

36

B-290 Operation Manual, Version I

a Tube to spraying nozzle
b Handle, pump bed
c Distilled water
d Flow meter valve
e Adjusting screw, pump bed

Fig. 5.12: Pump bed adjustment

1. Remove spraying nozzle from the Mini Sprayer B-290 and hold spraying nozzle into a suitable

container.

2. Insert tube a in to the pump bed.
3. Put tube on the suction side into distilled water c.
4. Close pump bed with handle b.
5. Unscrew adjusting screw e 1 turn.
6. Open the flow meter valve d to 40 mm.

NOTE
Distilled water
c is now flowing through the tubing without the feeding pump running.

7. Turn the adjusting screw e in until the water flow stops.
8. Turn adjusting screw ¼ further in.
9. Close flow meter valve to 0 mm.
10. Install spraying nozzle back into the Mini Sprayer B-290.

Choose the feeding tube according to the following table. For short processes, even unsuitable tubes
can be used, as swelling takes some time.

Table 5-1: Choosing the feeding tube

Solvent

Silicone tube

Tygon MH 2075

Tygon F 4040 A

Methanol

suitable

suitable

suitable

Ethanol

suitable

suitable

suitable

Acetone

unsuitable

suitable

unsuitable

Toluoene

unsuitable

unsuitable

unsuitable

Isopropanol

suitable

suitable

suitable

Chloroform

unsuitable

unsuitable

unsuitable

Dichloro methane

unsuitable

unsuitable

unsuitable

THF

unsuitable

unsuitable

unsuitable

Ethylacetate

unsuitable

suitable

unsuitable

Hexane

unsuitable

unsuitable

suitable