BUCHI Inert Loop B-295 User Manual
Page 36
5 Putting into operation
36
B-290 Operation Manual, Version I
a Tube to spraying nozzle
b Handle, pump bed
c Distilled water
d Flow meter valve
e Adjusting screw, pump bed
Fig. 5.12: Pump bed adjustment
1. Remove spraying nozzle from the Mini Sprayer B-290 and hold spraying nozzle into a suitable
container.
2. Insert tube a in to the pump bed.
3. Put tube on the suction side into distilled water c.
4. Close pump bed with handle b.
5. Unscrew adjusting screw e 1 turn.
6. Open the flow meter valve d to 40 mm.
NOTE
Distilled water c is now flowing through the tubing without the feeding pump running.
7. Turn the adjusting screw e in until the water flow stops.
8. Turn adjusting screw ¼ further in.
9. Close flow meter valve to 0 mm.
10. Install spraying nozzle back into the Mini Sprayer B-290.
Choose the feeding tube according to the following table. For short processes, even unsuitable tubes
can be used, as swelling takes some time.
Table 5-1: Choosing the feeding tube
Solvent
Silicone tube
Tygon MH 2075
Tygon F 4040 A
Methanol
suitable
suitable
suitable
Ethanol
suitable
suitable
suitable
Acetone
unsuitable
suitable
unsuitable
Toluoene
unsuitable
unsuitable
unsuitable
Isopropanol
suitable
suitable
suitable
Chloroform
unsuitable
unsuitable
unsuitable
Dichloro methane
unsuitable
unsuitable
unsuitable
THF
unsuitable
unsuitable
unsuitable
Ethylacetate
unsuitable
suitable
unsuitable
Hexane
unsuitable
unsuitable
suitable