PMP alarm codes
Error condition
Definition / Description
Required action
Alarm limit exceed. NIBP diastolic
HIGH.
Alarm limit exceeded. NIBP MAP
HIGH.
Alarm limit exceeded. NIBP systolic
HIGH.
Alarm limit exceeded. Pulse rate
HIGH.
Alarm limit exceeded SpO2 HIGH.
Alarm limit exceeded. Temperature
HIGH.
The alarm is at or has exceeded the
high alarm limit; alarm is over
range.
Change alarm limit or power device
off, and then on.
Alarm limit exceed. NIBP diastolic
LOW.
Alarm limit exceeded. NIBP MAP
LOW.
Alarm limit exceeded. NIBP systolic
LOW.
Alarm limit exceeded. Pulse rate
LOW.
Alarm limit exceeded SpO2 LOW.
Alarm limit exceeded. Temperature
LOW.
The alarm is at or has exceeded the
high alarm limit; alarm is over
range.
Change alarm limit or power device
off, and then on.
Searching for pulse signal.
SpO2 unknown.
No pulse signal.
No pulse is found.
Change alarm limit or power device
off, and then on.
Low battery 5 minutes remaining.
The battery is extremely low. Initial
alarm condition occurs when 5
minutes of battery power remain;
this is the second low battery
alarm.
Click the OK button to dismiss the
message.
Plug the device into the outlet to
dismiss the message.
Low battery 30 minutes remaining. The battery is extremely low. Initial
alarm condition occurs when 30
minutes of battery power remain;
this is the first low battery alarm.
Click the OK button to dismiss the
message.
Plug the device into the outlet to
dismiss the message.
5