Setup – Welch Allyn Connex Vital Signs Monitor PMP Tab Instructions - Quick Reference Guide User Manual

Page 16

background image

Part number

Description

Quantit
y

1/8 tube

1

Setup

1. Connect one end of a stereo jack plug to the PMP service test box and the opposite end to the

device.

2. Connect a USB cable to the PMP service test box; connect the opposite end to a PC.

3. Connect a mini B USB cable to the PMP service test box and the opposite end to the device

(comms port 1).

4. Connect the blood pressure hose to the NIBP module and the test volume.

Connect a piece of tubing to the squeeze bulb with bleeding screw; connect the opposite

end to the square fitting connector.

Connect another piece of tubing to the square fitting connector; connect the opposite end

to the calibrated pressure meter.

Connect another piece of tubing to the square fitting connector; connect the opposite end

to the test volume.

Connect the last piece of tubing to the square fitting connector; connect the opposite end

to the "Y" fitting on the two-tube blood pressure connector.

12 PMP calibration and verification setup