5. Connect the SpO2 tester.

If testing a Masimo sensor, plug the SpO2 sensor directly into the device.

If testing a Nellcor sensor, attach the Nellcor cable to the sensor and plug the opposite

end into the device.

6. Connect the Netgear Wireless-G WGR614 V9 or equivalent router.

Connect an ethernet cable from the PC to the router.

Connect another ethernet cable from the router to the device.

7. Plug in the 9600 Plus Calibration Tester. The tester takes approximately 20 minutes to heat to

the lowest setting. When testing several thermometers at all three temperatures, it is

recommended to test all probes at one Calibration Set Point Temperature before proceeding to

the next Calibration Set Point Temperature.

To expedite testing, start at the lowest Calibration Set Point Temperature. The 9600 Plus

Calibration Tester does not have an internal fan, which results in a longer cool down time than

warm up time. Refer to the 9600 Plus Calibration Tester Directions for Use manual for specific

information regarding the LCD window or the control buttons.

Choose the proper mains plug insert and slide it over the two prongs in the power

converter.

PMP calibration and verification setup 13