Error condition

Definition / Description

Required action

Dubious reading will be

accompanied by splat (*).

Retake NIBP.

Tissue contact lost.

Lost tissue contact while

attempting to acquire temperature

reading or reading was performed

with limited tissue contact

Click the OK button to dismiss the

message.
Retake NIBP.

Device is operating in battery

mode.

AC power cord has been

disconnected.

Click the OK button to dismiss the

message.
Plug the device into the outlet to

dismiss the message.

Radio card rebooting. Please wait. Device should not be shut down

while radio software upgrade is in

progress.

Click the OK button to dismiss the

message.
Radio card reboots and dismisses

the message.

Unable to save configuration to

USB.

There was a problem writing the

device configuration to USB flash

drive.

Click the OK button to dismiss the

message.

Unable to save configuration from

USB.

There was a problem reading the

device configuration from the USB

flash drive.

Click the OK button to dismiss the

message.

Powering down.

Device is performing normal

shutdown activities.

System error; disconnect patient

cables, power down, and restart.

System failure: fault is detected

that could adversely affect the

product.

User must turn the VSM off using

the On/Off button.

Use screen slide to unlock.

User selects area outside of display

unlock slide control.

Click the OK button to dismiss the

message.
Touch screen unlock control

dismisses the message.

Demonstration mode.

DEMO mode is active.

Message will be provided until

normal operation is selected.

No data to save.

User attempts to save information,

but no information is available.

Click the OK button to dismiss the

message.

Unable to automatically send

readings.

Automatic send with manual save

configured, but there is no

connectivity on command.

Click the OK button to dismiss the

message.

Data did not transfer.

Data transfer unsuccessful with

send.

Click the OK button to dismiss the

message.

Data transfer successful.

Successful data transfer with

automatic or manual save

Click the OK button to dismiss the

message.

Patient ID required to save data.

Device requires Patient ID to

manually save.

Click the OK button to dismiss the

message.

PMP alarm codes 9