Error condition
Definition / Description
Required action
Dubious reading will be
accompanied by splat (*).
Retake NIBP.
Tissue contact lost.
Lost tissue contact while
attempting to acquire temperature
reading or reading was performed
with limited tissue contact
Click the OK button to dismiss the
message.
Retake NIBP.
Device is operating in battery
mode.
AC power cord has been
disconnected.
Click the OK button to dismiss the
message.
Plug the device into the outlet to
dismiss the message.
Radio card rebooting. Please wait. Device should not be shut down
while radio software upgrade is in
progress.
Click the OK button to dismiss the
message.
Radio card reboots and dismisses
the message.
Unable to save configuration to
USB.
There was a problem writing the
device configuration to USB flash
drive.
Click the OK button to dismiss the
message.
Unable to save configuration from
USB.
There was a problem reading the
device configuration from the USB
flash drive.
Click the OK button to dismiss the
message.
Powering down.
Device is performing normal
shutdown activities.
System error; disconnect patient
cables, power down, and restart.
System failure: fault is detected
that could adversely affect the
product.
User must turn the VSM off using
the On/Off button.
Use screen slide to unlock.
User selects area outside of display
unlock slide control.
Click the OK button to dismiss the
message.
Touch screen unlock control
dismisses the message.
Demonstration mode.
DEMO mode is active.
Message will be provided until
normal operation is selected.
No data to save.
User attempts to save information,
but no information is available.
Click the OK button to dismiss the
message.
Unable to automatically send
readings.
Automatic send with manual save
configured, but there is no
connectivity on command.
Click the OK button to dismiss the
message.
Data did not transfer.
Data transfer unsuccessful with
send.
Click the OK button to dismiss the
message.
Data transfer successful.
Successful data transfer with
automatic or manual save
Click the OK button to dismiss the
message.
Patient ID required to save data.
Device requires Patient ID to
manually save.
Click the OK button to dismiss the
message.
PMP alarm codes 9