Error condition
Definition / Description
Required action
failure in power supply, a printer
that does not identify itself
correctly, or a printer that does not
enumerate.
Radio not functional. Call for
service.
This message could be caused by a
hardware failure, wrong software
installed in radio or a disconnected
radio.
Reconnect USB cable. Check
wireless icon. Reposition radio.
Radio error. Power down and
restart.
Radio card is inoperable; unable to
establish communications between
VSM and radio or failed read/write
operation with radio.
Check wireless icon. Restart radio.
Unable to establish network
connections. Radio out of network
range.
Radio was communicating but is
out of range or is unable to
continue communications with the
access point.
Check wireless icon. Reposition
radio.
Unable to establish network
connections. Call for service.
Unable to establish
communications between radio and
DHCP server (network).
Communication cannot proceed
without an IP from the DHCP
server.
Radio software upgrade failed.
Radio software upgrade failed to
complete successfully.
Unable to load configuration.
Unable to load non-critical
configuration item(s) from NAND
Flash.
Device will use engineering and
factory defaults.
Functional error. Call for service.
Unable to load critical configuration
item(s) from NAND Flash.
Device should not be used and sent
for repair.
External device not recognized.
An unrecognized external device
has been connected.
USB communication failure. Call for
service.
This message could be caused
because an internal or external
device connected but failed
enumeration or an overcurrent
condition on internal or external
port could have occurred.
Disable the missing functionality
with warning to user. Recovery
from error may be possible with
power cycle.
User cancelled NIBP reading.
Blood pressure reading was
cancelled by the user.
Click the OK button to dismiss the
message.
Retake NIBP.
Tube type does not match device
configuration.
The tube connected to the NIBP
sensor does not match the VSM
configuration (tube type, patient
mode and algorithm).
Click the OK button to dismiss the
message.
Retake NIBP.
Excessive patient movement.
NIBP reading is available, yet not
precise due to movement, noise.
Click the OK button to dismiss the
message.
8 PMP alarm codes