Error condition

Definition / Description

Required action

failure in power supply, a printer

that does not identify itself

correctly, or a printer that does not

enumerate.

Radio not functional. Call for

service.

This message could be caused by a

hardware failure, wrong software

installed in radio or a disconnected

radio.

Reconnect USB cable. Check

wireless icon. Reposition radio.

Radio error. Power down and

restart.

Radio card is inoperable; unable to

establish communications between

VSM and radio or failed read/write

operation with radio.

Check wireless icon. Restart radio.

Unable to establish network

connections. Radio out of network

range.

Radio was communicating but is

out of range or is unable to

continue communications with the

access point.

Check wireless icon. Reposition

radio.

Unable to establish network

connections. Call for service.

Unable to establish

communications between radio and

DHCP server (network).
Communication cannot proceed

without an IP from the DHCP

server.

Radio software upgrade failed.

Radio software upgrade failed to

complete successfully.

Unable to load configuration.

Unable to load non-critical

configuration item(s) from NAND

Flash.

Device will use engineering and

factory defaults.

Functional error. Call for service.

Unable to load critical configuration

item(s) from NAND Flash.

Device should not be used and sent

for repair.

External device not recognized.

An unrecognized external device

has been connected.

USB communication failure. Call for

service.

This message could be caused

because an internal or external

device connected but failed

enumeration or an overcurrent

condition on internal or external

port could have occurred.

Disable the missing functionality

with warning to user. Recovery

from error may be possible with

power cycle.

User cancelled NIBP reading.

Blood pressure reading was

cancelled by the user.

Click the OK button to dismiss the

message.
Retake NIBP.

Tube type does not match device

configuration.

The tube connected to the NIBP

sensor does not match the VSM

configuration (tube type, patient

mode and algorithm).

Click the OK button to dismiss the

message.
Retake NIBP.

Excessive patient movement.

NIBP reading is available, yet not

precise due to movement, noise.

Click the OK button to dismiss the

message.

8 PMP alarm codes