Place the 9600 Plus Calibration Tester on a level surface away from sunlight, drafts, and

other sources of heat or cold.

Observe the Set Point Mode in the upper left hand corner of the LCD display. If the unit

displays a "D", it is in Default Mode and will heat to the lowest set point temperature. If you

do not want to conduct testing at this set point temperature, press and hold the

Temperature Selection button to select the desired setting. The temperature display will

flash before staying on continuously to indicate the 9600 Plus Calibration Tester has

stabilized and is ready for use.

To change the calibration set point temperature, scroll from one set point to the next,

press and hold the Temperature Selection button until a beep is heard. The newly

selected set point appears in the upper left corner of the LCD display. The device’s

current temperature is displayed, starts to flash, and continues flashing until the cavity

reaches the equilibrium at the new set point.

8. Remove the probe from the probe well and clean it with a 70% isopropyl alcohol solution, a

10% chlorine bleach solution, or a non-staining disinfectant. Let the probe air dry. Do not apply

a probe cover.

PMP calibration and verification setup 15