Welch Allyn Connex Vital Signs Monitor PMP Tab Instructions - Quick Reference Guide User Manual
Page 10

Error condition
Definition / Description
Required action
Powering down. Call for service.
This message could be caused by
various internal errors, power
manager, and battery faults.
The device will auto power off.
Battery is absent or faulty.
Battery is absent or faulty.
Plug the device into the outlet to
dismiss the message.
NIBP air leak; check cuff and
tubing connections.
An air leak has been detected.
Take corrective action; retake
measurement.
NIBP not functional. Call for
service.
This message could be caused by
various internal errors, a calibration
error, or ambient temperature or
module is out of range.
Retake measurement.
Unable to determine NIBP; check
cuff and tubing connections.
This message could be caused
because pressure exceeded the
maximum limit for the patient mode,
too much motion artifact to
calculate a reading, the system
dumped too soon, not enough data
could be obtained or the device
needed to re-inflate after it was
safe.
Limit patient movement; retake
measurement.
Unable to determine NIBP. Check
connections and tubing for kinks.
This message could be caused
because pressure exceeded the
maximum limit for the patient mode
or because pressure could not be
bled.
Retake measurement.
Incorrect NIBP cuff size. Check
patient type.
The patient is outside of the
module's capability to measure
NIBP.
Retake measurement.
Inflation too quick. Check NIBP cuff
and tubing connections.
Inflation was too quick; check for a
kinked hose.
Take corrective action; retake
measurement.
Unable to determine NIBP. Check
inflation settings.
Reading couldn’t be obtained
because a high enough inflation
couldn’t be reached.
Retake measurement.
SpO2 not functional. Call for
service.
This message could be caused by
various internal errors or a non-
functional SpO2 module, or SpO2
sensor may need to be replaced.
Attach SpO2 to monitor.
Sensor is not attached.
Check connections.
Replace SpO2 sensor.
SpO2 Sensor needs to be replaced.
This message may be caused by
BAD_PROBE error.
Set date and time.
Either date or time not set at all or
data has not been set.
Set date and time.
6 PMP alarm codes