background image

Error condition

Definition / Description

Required action

Ambient temperature outside

operating range. Retry

measurement.

Device outside operating

temperature range.

An alarm will sound, but you may

continue to take measurements.

Maximum number of patient

records saved. Oldest record

overwritten.

The maximum number of test

records are stored in memory and

save is attempted.

An alarm will sound and

automatically overwrite oldest

record with new record. You can

prevent alarm from activating for

subsequent saves by deleting them

patient records from the Review

tab.

Data did not transfer.

Data transfer unsuccessful with

save.

Try to transfer data again.

Unable to save.

Unable to save data/readings to

device.

Try to save data again.

Unable to access patient

information.

There was an error reading in the

patient list or the patient records

from the flash file system on

startup.

Connect temperature probe.

No probe is connected.

Connect a probe and try again. If a

probe is connected, replace the

probe. If problem persists, replace

device.

Insert correct color coded probe

well.

Missing Probe well.

Insert probe well.

Replace temperature probe.

This message could be caused by

various probe errors.

Reconnect temperature probe.

Temperature not functional. Call for

service.

This message could be caused by

internal or messaging errors.

Have module serviced.

Retry temperature measurement.

This message could be caused by

user settings, heater, or various

error log issues.

Try another predict cycle. If

problem persists, replace probe.

Low battery plug into outlet.

The monitor’s battery voltage is too

low to support printing.

Connect the AC power adapter to

recharge the battery.

Printer door is open. Close to

continue.

Printer cover is open.

Close the printer cover.

Printer not functional. Call for

service.

Printer cover may be open or the

detection switch has

malfunctioned.

Close the printer cover or have

printer serviced.

Out of paper.

Printer has run out of paper.

Replace the paper.

Printer too hot. Wait to retry.

Print head has overheated.

Wait for the print head to cool

down.

Printer not functional. Call for

service.

This message could be caused by a

broken printer motor, hardware

Have printer serviced.