Error condition
Definition / Description
Required action
Ambient temperature outside
operating range. Retry
measurement.
Device outside operating
temperature range.
An alarm will sound, but you may
continue to take measurements.
Maximum number of patient
records saved. Oldest record
overwritten.
The maximum number of test
records are stored in memory and
save is attempted.
An alarm will sound and
automatically overwrite oldest
record with new record. You can
prevent alarm from activating for
subsequent saves by deleting them
patient records from the Review
tab.
Data did not transfer.
Data transfer unsuccessful with
save.
Try to transfer data again.
Unable to save.
Unable to save data/readings to
device.
Try to save data again.
Unable to access patient
information.
There was an error reading in the
patient list or the patient records
from the flash file system on
startup.
Connect temperature probe.
No probe is connected.
Connect a probe and try again. If a
probe is connected, replace the
probe. If problem persists, replace
device.
Insert correct color coded probe
well.
Missing Probe well.
Insert probe well.
Replace temperature probe.
This message could be caused by
various probe errors.
Reconnect temperature probe.
Temperature not functional. Call for
service.
This message could be caused by
internal or messaging errors.
Have module serviced.
Retry temperature measurement.
This message could be caused by
user settings, heater, or various
error log issues.
Try another predict cycle. If
problem persists, replace probe.
Low battery plug into outlet.
The monitor’s battery voltage is too
low to support printing.
Connect the AC power adapter to
recharge the battery.
Printer door is open. Close to
continue.
Printer cover is open.
Close the printer cover.
Printer not functional. Call for
service.
Printer cover may be open or the
detection switch has
malfunctioned.
Close the printer cover or have
printer serviced.
Out of paper.
Printer has run out of paper.
Replace the paper.
Printer too hot. Wait to retry.
Print head has overheated.
Wait for the print head to cool
down.
Printer not functional. Call for
service.
This message could be caused by a
broken printer motor, hardware
Have printer serviced.
PMP alarm codes 7