beautypg.com

Welch Allyn 408 Micropaq Monitor - User Manual User Manual

Page 73

background image

Directions for use

Chapter 7 Compliance

69

Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the
Model 4XX Series Monitor

The Model 4XX Series Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the monitor can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the monitor as
recommended below, according to the maximum output power of the communications equipment.

Rated Maximum Output Power of
Transmitter (W)

Separation Distance (m) According to Frequency of Transmitter

150 kHz to 80 MHz

d = 1.2

80 MHz to 800 MHz

d = 1.2

800 MHz to 2.5 GHz

d = 2.3

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (w) according to the transmitter manufacturer.

NOTE 1

At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption