beautypg.com

Welch Allyn 408 Micropaq Monitor - User Manual User Manual

Page 18

background image

14

Chapter 1 General information

Welch Allyn Micropaq Monitor

ECG SCALE...

Change the scale of the ECG waveform. If two waveforms are
displayed, both have the same scale.

1 WAVEFORM

There are four possible ECG waveform display selections:

1 WAVEFORM

the default selection

2 WAVEFORMS

5 SECONDS

FULL SCREEN

Pressing

changes to the next selection. This change does not

take effect until after you exit the Main Menu. See

”Display”

on

page 12 for descriptions.

LIMITS...

Enter the Alarm Limits Menu (

”Customize patient alarm limits at

the monitor”

on page 37) and change alarm limits.

SYSTEM
INFORMATION

Display information about the network connection and SpO

2

module.

SERVICE MENU Enter Service Mode for a demonstration mode (Demo, see

”Demonstration mode”

on page 17) or service functions for

technicians. Service Mode is not available if any cables are
plugged in.

Note

To restrict access to the Main Menu, a Menu Lock option can be configured for
the monitor at the Acuity Central Station. When the Menu Lock is enabled, the
operator must press and hold down

and

for two seconds to gain access

to the Main Menu. The Menu Lock is disabled if the monitor loses
communication with Acuity.

This manual is related to the following products: