Patient alarm and equipment alert specifications

Characteristic

Specification

Visual Alarm Indicator at the monitor

Flashing GREEN LED
Flashing RED LED
Continuously ON RED LED
Flashing YELLOW LED
Continuously ON YELLOW LED

Normal operation
Patient alarm
Patient alarms are silenced
An equipment alert or not connected to the network
Equipment alert suspended for 90 seconds at Acuity or low battery alert
acknowledged (dismissed)

Audio Tone Locations

Monitor
Acuity Central Station (when connected)

Audio Tone Frequency

2900 Hz

Life-Threatening Arrhythmia Alarm
Tone Pattern
Patient Alarm Tone Pattern
Equipment Alert Tone Pattern

1 second on, 1 second off
1second on, 2 seconds off
1second on, 4 seconds off

Audio Tone Volume

The monitor audio tone volume is configured by the Acuity System to High,
Low, or Off. The monitor can be configured with separate audio tone volume
settings for when it is connected to an Acuity System and when it is not.

Limits

Setable on all parameters

Alarm Control

Automatic preset or manual settings

Alarm Priority

Highest priority: Patient alarms
Lowest priority: Equipment alerts

Alarm on Tachycardias

Most tachycardias will alarm in less than 8 seconds. These include AAMI
3.1.2.1 part f. waveforms. Certain multifocal tachycardias may initially alarm
as "low rate."

Alarm Holdoff Time Period

a

HR = 3 seconds
% SpO

2

, PR = 10 seconds

Acuity-Configurable Audio Alarm Delay
at the monitor

When a monitor is connected to an Acuity System, the audio alarms at the
monitor can be delayed up to 4 minutes and 15 seconds. The delay time is
selected in Acuity software at the time of Acuity installation. Visual alarm
indications are not delayed.

Patient Alarm Tone Silence from the
monitor or Suspend from Acuity

The monitor LED is continuously ON RED and the audio tone is silenced for 90
seconds (non-adjustable).

If original alarm was silenced from the monitor, new patient alarms or
equipment alerts break the silence at the monitor, but only life-threatening
arrhythmia alarms break the silence at Acuity. If original alarm suspended at
Acuity, only life-threatening arrhythmia alarms break the silence at the
monitor and Acuity.

Equipment Alert Acknowledge from the
monitor

The LED returns to the pre-alert state (except Low Battery remains
continuously ON YELLOW) and the auditory tone is dismissed.

Equipment Alert Suspend from Acuity

The LED is continuously ON YELLOW and the audio tone is silenced for 90
seconds (non-adjustable).

Patient Alarm Tone Reset from the
monitor or Resume from Acuity

For a patient alarm tone that has been silenced, resets the tone.

Patient Out of Range; Transmitter
Failure

An equipment alert is generated whenever the monitor fails to communicate
to an Acuity System after a connection has been successfully established. In
addition, the “No Acuity” icon is displayed on the monitor display.