Reference
Operating settings
The following table lists all of the monitor settings and the default settings.
Parameter
Set at
Monitor
Set at Acuity
Previous
Setting
Retained at
Monitor at
Power-Up
Monitor Default
Setting
For Each
Patient
a
a.
Set by clinician at Acuity Central Station.
For Entire
Acuity
Unit
b
b.
Set by Acuity System Administrator during system installation.
Patient Mode
Adult (age 13 years and older)
Pediatric (age greater than 28
days of age or more than 44
weeks gestation up to 12
years)
No
Yes
No
Yes
Adult
ECG screen mode (Single, Dual, 5
Sec, or Full Screen)
Yes No
Yes
Yes
Single
ECG 1 Lead Selection
Yes
Yes
Yes
Yes
II
ECG 2 Lead Selection
Yes
Yes
Yes
Yes
V (or III if no V lead)
ECG Size (Scale)
Yes
No
No
Yes
1 mV/cm
Language No
No
Yes
Yes
English
Mains Filter (off, 50, or 60 Hz)
No
No
Yes
Yes
60
Vital Signs Alarm Volume
(high, low, or off)
With Acuity Connection
Without Acuity Connection
No
No
No
No
Yes
Yes
No
Yes
off
high
Equipment Alert Volume
(high, low, or off)
With Acuity Connection
Without Acuity Connection
No
No
No
No
Yes
Yes
No
Yes
off
low
HR/PR Alarm Limits
(Lower, Upper)
Yes
Yes
Yes
Yes
Adult: 50, 120 bpm
Ped: 50, 150 bpm
SpO
2
Alarm Limits
c
(Lower, Upper)
c.
SpO
2
alarm limit range depends on the software version of the Acuity System to which the monitor is connected. (See
Yes
Yes
Yes
Yes
Adult: 90, 100%
Ped: 90, 100%
Regulatory settings
(U.S., Europe, Japan)
No
No
Yes
Yes
U.S.
Pacer Detection Enable
No
Yes
No
Yes
On
Menu Lockout
No
No
Yes
No
Off
Display Backlight Timeout
No
No
Yes
Yes
120 seconds