Chapter 3 Alarms & alerts

Welch Allyn Micropaq Monitor

Respond to a patient alarm at monitor

When a patient alarm occurs, the monitor produces an audible tone (if audible tones are
enabled). Life-threatening arrhythmia alarms beep at a faster pace than other vital sign
alarms (see

“Patient alarm and equipment alert specifications”

on page 58). The monitor

also displays a message similar to the following:

1.

Check the patient and provide appropriate care.

2.

To silence the alarm tone at the monitor and the Acuity Central Station for 90
seconds, press

.

While the alarm tone is silenced, visual alarm indications continue, and the red
alarm indicator on the monitor changes from a flashing display to a continuous
display.

If the alarm condition still exists after 90 seconds, the alarm tone resumes.

If you suspend an alarm at Acuity, only life-threatening arrhythmia alarms interrupt
the silence period at the monitor and Acuity.

To access the Main Menu during silencing, press

.

Press to silence the alarm tone.

Red LED flashes

during alarm.

After the alarm is

silenced (or

suspended at

Acuity), the red

LED is on

continuously.

Flashing numerics

indicate patient

alarm.

Note

If you silence an alarm at the monitor and another patient alarm or an equipment
alert occurs during the silence period, the tone resumes at the monitor. At
Acuity, only life-threatening arrhythmia alarms interrupt the silence period.