Physical specifications battery specifications – Welch Allyn 408 Micropaq Monitor - User Manual User Manual
Page 64

60
Chapter 6 Reference
Welch Allyn Micropaq Monitor
Physical specifications
Battery specifications
Protection classifications, all configurations
Characteristic
Specification
Type of Protection against Electric
Shock—Monitor Type: CF
Battery operation only
Battery must be recharged in separate battery charger.
IEC EN 60601-1, 2nd Edition
Degree of Protection Against Electric
Shock, for Parts Applied to Patients
See monitor labels. CF defibrillator protected.
IEC EN 60601-1, 2nd Edition
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Method of Disinfection
Not suitable for autoclaving (see cleaning instructions on
monitor, batteries, battery charger, and accessories”
Flammable Anesthetics
Not suitable for use with flammable anesthetics.
Height
7.80 in (19.8 cm)
Width
3.50 in (8.9 cm)
Depth
1.96 in (4.9 cm)
Weight (including battery)
Model 406
Model 408
17.0 oz (0.48 kg)
18.6 oz (0.53 kg)
Characteristic
Specifications
Reorder Number
Lithium Ion Battery
2EA Active A
008-0647-XX
Battery Type
Rechargeable, Lithium Ion
Battery Capacity
2 cells, 7.4 V (nominal), 8.4 V (charging), 1800 mA-hr
Battery Weight
4.5 oz (0.13 kg)
Battery Charger
External device
Battery Fuse Rating
5 A, 125 V (not user-accessible)
Note: Internal electronic overload circuitry is used as
the primary method of protection.
Operating Times on Battery
a
a.
Battery operating times based on these conditions: new fully-charged battery operating at 25° C,
the monitor connected to Acuity, eight patient alarms per hour, minimal motion artifact.
Model 406: 25 hrs
Model 408: 10 hrs
Battery Recharge Time
4 hours at 25° C (typical)
Battery Lifetime
300 charge/discharge cycles to 70% of original
capacity (typical)
Note
The following factors may reduce battery operating time:
•
Amount of time not connected to Acuity.
•
Frequency and duration of alarms and alerts.
•
Amount of operator activity using monitor keys (activates display).
•
Age of battery.
•
Amount of motion artifact during SpO
2
monitoring.
WARNING The monitor battery is Lithium Ion. Do not incinerate, submerge,
crush, disassemble, or autoclave. If a battery has been submerged in liquid,
discard the battery properly; do not try to recharge or reuse the battery. Do not
short the battery terminals. Do not try to connect the battery to any device
except the monitor or the battery charger. Do not expose to high temperature
(above 60° C or 140° F). Use only the specified monitor battery charger.