Operation 37

At the time of writing, the setpoints, hysteresis values are
saved to the RAM and will be lost upon the next power-off; to
store them permanently to the EEPROM so that they are
maintained at power-on, the ‘99’ command of the Command
Register must be sent.

REGISTER

DESCRIPTION

Saving to EEPROM ACCESS

40001

Firmware version

-

R

40002

Type of instrument

-

R

40003

Year of Production

-

R

40004

Serial Number

-

R

40005

Active program

-

R

40006

Command Register

NO

W

40007

Status Register

-

R

40008

Gross Weight H

-

R

40009

Gross Weight L

-

R

40010

Net Weight H

-

R

40011

Net Weight L

-

R

40012

Peak Weight H

-

R

40013

Peak Weight L

-

R

40014

Divisions and Units of measure

-

R

40015

Coefficient H

R

40016

Coefficient L

R

40017

SETPOINT 1 H

Only after command

‘99’ of the COMMAND

REGISTER

R/W

40018

SETPOINT 1 L

R/W

40019

SETPOINT 2 H

R/W

40020

SETPOINT 2 L

R/W

40021

HYSTERESIS 1 H

R/W

40022

HYSTERESIS 1 L

R/W

40023

HYSTERESIS 2 H

40024

HYSTERESIS 2 L

40025

INPUTS

-

R

40026

OUTPUTS

NO

R/W

40037

Test weight for calibration H

Use with command

‘101’ of the

COMMAND REGISTER

R/W

40038

Test weight for calibration L

R/W

40043

Weight value corresponding to
ZERO of the analog output H

Only after command

‘99’ of the

Command Register.

R/W

40044

Weight value corresponding to
ZERO of the analog output L

R/W

40045

Weight value corresponding to
Full Scale of the analog output H

R/W

40046

Weight value corresponding to
Full Scale of the analog output L

R/W

Important