Rice Lake SCT-10 User Manual
Page 41

Operation 37
At the time of writing, the setpoints, hysteresis values are
saved to the RAM and will be lost upon the next power-off; to
store them permanently to the EEPROM so that they are
maintained at power-on, the ‘99’ command of the Command
Register must be sent.
REGISTER
DESCRIPTION
Saving to EEPROM ACCESS
40001
Firmware version
-
R
40002
Type of instrument
-
R
40003
Year of Production
-
R
40004
Serial Number
-
R
40005
Active program
-
R
40006
Command Register
NO
W
40007
Status Register
-
R
40008
Gross Weight H
-
R
40009
Gross Weight L
-
R
40010
Net Weight H
-
R
40011
Net Weight L
-
R
40012
Peak Weight H
-
R
40013
Peak Weight L
-
R
40014
Divisions and Units of measure
-
R
40015
Coefficient H
R
40016
Coefficient L
R
40017
SETPOINT 1 H
Only after command
‘99’ of the COMMAND
REGISTER
R/W
40018
SETPOINT 1 L
R/W
40019
SETPOINT 2 H
R/W
40020
SETPOINT 2 L
R/W
40021
HYSTERESIS 1 H
R/W
40022
HYSTERESIS 1 L
R/W
40023
HYSTERESIS 2 H
40024
HYSTERESIS 2 L
40025
INPUTS
-
R
40026
OUTPUTS
NO
R/W
40037
Test weight for calibration H
Use with command
‘101’ of the
COMMAND REGISTER
R/W
40038
Test weight for calibration L
R/W
40043
Weight value corresponding to
ZERO of the analog output H
Only after command
‘99’ of the
Command Register.
R/W
40044
Weight value corresponding to
ZERO of the analog output L
R/W
40045
Weight value corresponding to
Full Scale of the analog output H
R/W
40046
Weight value corresponding to
Full Scale of the analog output L
R/W
Important