14 SCT Weight Transmitter Operator’s Manual
2.1.1 Theoretical Calibration
This function allows load cell rated values to be set.
To perform the theoretical calibration set FS-tED, SEnS lb and dI UI S in
sequence:
When entering a menu, the LED’s will begin scrolling, when
selection is made and confirmed the LED’s will be off.
1. Press and hold
, then press
, CALIb will be displayed.
2. Press ,
FS-tEO is displayed. Press
.
3. Press
or
until total load cell capacity (system full scale) is
displayed, press
.
4. Press
or
until SEnS lb is displayed, press
.
5. Press
or
until desired load cell mV/V is displayed,
press .
6. Press
or
until dI UI S is displayed, press
.
7. Press
or
until desired display division size is displayed,
press .
8. This complete the Theoretical Calibration, press
twice to exit set-
up menu or continue to Section 2.1.2.
By modifying the theoretical full scale, the sensitivity or divisions,
the weight (span) calibration is cancelled and the theoretical
calibration only is considered valid.
If the theoretical full scale and the recalculated full scale in weight
(span) calibration (see Section 2.1.5) are equal, this means that the
calibration currently in use is theoretical; if they are different, the
calibration in use is the weight (span) calibration based on test
weights.
By modifying the theoretical full scale, the sensitivity or divisions
and all the system’s parameters containing a weight value will be
set to default values (setpoints, hysteresis, etc.).
2.1.2 Maximum Capacity (NASS )
Maximum capacity (live load/product) that can be displayed. When the weight
exceeds this value by 9 divisions the following is displayed ‘------’, indicating
overload. To disable this function, set to 0.