28 SCT Weight Transmitter Operator’s Manual
2.7
Test
1. Press and hold
, then press
. CALIb will be displayed.
2. Press
or
until tEst is displayed, press
. The currently
programmed value is displayed.
3. Press
or
until desired parameter is displayed, press
.
4. For In and NU-CEL, current reading is displayed, press
.
For Out, press
until corresponding value of the out you want to
change is flashing, press
to change the value, press
.
5. Press
twice to exit set-up menu.
In 1
In 2
nE-LO *
(In 2
default)
(NET/GROSS): by closing this input for less than one
second, it performs a SEMI-AUTOMATIC TARE and the
display will show the net weight. To display the gross
weight again, hold the NET/GROSS input closed for 3
seconds.
2ErO *
(In 1
default)
By closing the input for less than one second, the
weight is set to zero (see Section 3.3 “Semi-Automatic
Zero (Weight Zero-setting For Small Variations)” on
page 33).
PEAH
With the input closed the maximum weight value
reached remains on display. Opening the input the
current weight is displayed.
PLC
Closing the input no operation is performed, the input
status may however be read remotely by way of the
communication protocol.
COntin
Closing the input for less than one second the weight is
transmitted via the serial connection according to the
fast continuous transmission protocol one time only
(only if contin is set in the item serial).
COEFF
When the input is closed the weight is displayed based
on the set coefficient (see Section 2.1.6 “Setting Units
of Measure” on page 30 and Section 2.1.7 “Display
Coefficient” on page 31), otherwise the weight is
displayed.
*
- indicates default value.
Parameter
Choices
Description
Table 2-8. Output and Input Menu
MENU
0
ESC
TARE
MENU
TARE
MENU
MENU
TARE
MENU
0
ESC