3 zero setting, 4 zero value manual entry, 3 zero setting 2.1.4 zero value manual entry – Rice Lake SCT-10 User Manual
Page 19

Configuration 15
2. Press ,
FS-tEO is displayed.
3. Press
or
until
NASS is displayed, press
. LED’s will
begin scrolling.
4. Press
or
until desired capacity is displayed, press
.
5. Press
twice to exit set-up menu.
2.1.3 Zero Setting
Perform this procedure after having set the “Theoretical Calibration” on page 14.
This menu may also be accessed directly from the weight display,
press and hold
for 3 seconds.
1. Press and hold
, then press
. CALIb will be displayed.
2. Press ,
FS-tEO is displayed.
3. Press
or
until ZErO is displayed, press
.
4. The weight value to be set to zero is displayed. In this phase all of the
LEDs are flashing. Press
, the weight is set to zero (the value is
stored to the permanent memory).
5. Press
twice to exit set-up menu.
Press
to display the accumulated deadload that has been
zeroed off by the instrument, displaying the sum of all of the
previous zero settings.
2.1.4 Zero Value Manual Entry
Perform this procedure only if it is not possible to reset the
weighed structure tare, for example because it contains
product that can not be unloaded.
Set in this parameter the estimated zero value.
1. Press and hold
, then press
. CALIb will be displayed.
2. Press ,
FS-tEO is displayed.
3. Press
or
until
INP 0 is displayed, press
. LED’s will
begin scrolling.
4. Press
or
until desired dead load is displayed, press
.
5. Press
twice to exit set-up menu.
MENU
TARE
MENU
TARE
MENU
0
ESC
Note
0
ESC
MENU
0
ESC
MENU
TARE
MENU
MENU
0
ESC
Note
TARE
Important
MENU
0
ESC
MENU
TARE
MENU
TARE
MENU
0
ESC