6 setting units of measure, 7 display coefficient – Rice Lake SCT-10 User Manual
Page 21

Configuration 17
The procedure ends by pressing
or after entering the fifth value; at this
point it will no longer be possible to change the calibration value, but only to
perform a new weight (span) calibration. To perform a new calibration, should
return to the weight display and then re-entering into the calibration menu.
By pressing
after having confirmed the test weight that has been set, the full
scale appears, recalculated according to the value of the maximum test weight
entered and making reference to the cell sensitivity set in the theoretical
calibration (SEnSI b).
2.1.6 Setting Units of Measure
1. Press and hold
, then press
. CALIb will be displayed.
2. Press ,
FS-tEO is displayed.
3. Press
or
until unIt is displayed, press
.
4. Press
or
until desired unit is displayed, press
.
5. Press
twice to exit set-up menu.
* Indicates it is possible to set the display coefficient. To use COEFF it is
necessary to enable it, closing the COEFF input. See Section 2.1.7 “Display
If the print function is enabled, the symbol of the selected unit of
measure will be printed after the measured value.
2.1.7 Display Coefficient
By setting the coefficient the display is changed accordingly.
If one of the inputs is set to COEFF mode (see Section 2.6 “Outputs And Inputs
Configuration” on page 26) when the input is closed the value will be displayed
modified according to the coefficient; when the input is opened the standard
weight display will be restored.
1. Press and hold
, then press
. CALIb will be displayed.
2. Press ,
FS-tEO is displayed.
3. Press
or
until COEFF is displayed, press
. LED’s will
begin scrolling.
HI LOG
kilograms
bAr
bar*
G
grams
AtN
atmospheres*
t
tons
PI ECE
pieces*
Lb
pounds*
nEU-N
newton metres*
nEUton
newton*
HI LO-N
kikgram metres*
LI tre
litres*
OtHEr
other generic units of measure
not included in list*
0
ESC
MENU
0
ESC
MENU
TARE
MENU
TARE
MENU
0
ESC
Note
MENU
0
ESC
MENU
TARE
MENU