9 xbpi calibration with default weight, Xbpi calibration with default weight, Commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 80: Pr 5220 instrument manual

Commissioning
PR 5220 Instrument Manual
EN-80
Sartorius
In this example, a weight of 5000 g is put onto the scale.
The following window is displayed after applying the weight:
Weighingpoint/
xBPI-Scale Setup
Calibration status
Difference display
Cal-Delta
0.1 g
--
5000.06 g The weight is displayed in high-resolution (10x).
Select [Accept].
Accept
ResError
Abort
The data are saved and the instrument generates a corresponding message:
Weighingpoint/
xBPI-Scale Setup
Calibration status
complete
Net=Grs
5000.0 g
--
5000.01 g The weight is displayed in high-resolution (10x).
Leave with
.
Accept
ResError
Abort
4.5.9
xBPI Calibration with Default Weight
Prerequisites:
-
The xBPI protocol has been selected (see Chapter 4.5.1).
-
The weighing point ‚xBPI-scale’ has been selected (see Chapter 4.5.4).
-
The platform configuration has been executed (see Chapter 4.5.3).
-
The setting in the following menu has been done: [Weighingpoint A-xBPI-Scale]-[Setup]:
[Configuration]-[Confirming adjust.]: [manual] (see Chapter 4.5.5).
-
The communication between instrument and platform is active.
Weighingpoint/
xBPI-Scale Setup
Calibration
Open with .
Configuration
Select group of specification
Show device info
Weighingpoint/
xBPI-Scale Setup
Adjust with default weight
Select with
and enter with
.
Adjust with intern weight
Linearity
Default
User