4 weighing electronics, 1 load cells, 2 principle – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 161: 3 accuracy and stability, 4 sensitivity, Weighing electronics, Load cells, Principle, Accuracy and stability, Sensitivity

PR 5220 Instrument Manual
Specifications
Sartorius
EN-161
14.4
Weighing Electronics
The weighing electronics are suitable for connection of strain-gauge load cells.
14.4.1
Load Cells
Load cell type
Strain gauge, 6 or 4-wire connection possible
Supply voltage
U= 12V DC for I
max
= 160 mA
for up to 8 load cells of 650
Ω each
or 4 load cells of 350
Ω each
Sense voltage monitoring
Sense voltage below +4VDC -- -4VDC is detected; can be switched off
Max. load
≥75 Ω
Cable length
Max. 500 m
14.4.2
Principle
Principle
DC voltage, delta-sigma converter, ratiometric to the load cell supply voltage
Conversion/measurement time
5, 10, 20, 40, 80, 160, 320, 640, 960, 1200, 1600 ms
Digital filter
Selectable, active 4th order (low-pass),
Characteristic: Bessel, aperiodic, Butterworth, Tschebyscheff
Configurable cut-off frequency
14.4.3
Accuracy and Stability
Accuracy class
≤10000 e (Cl. III) acc. to OIML R76 / EN 45501
Min. measuring signal (OIML)
0.5µV/e, i.e. 3 mV for 6000 e, 5 mV for 10000 e
Linearity error:
<0.002 %
Zero stability error (TK
0
)
<0.02 µV/K RTI <= 0.004%/10K at 1mV/V
Span stability error (TK
Spn
)
< ±2 ppm/K
14.4.4
Sensitivity
Sensitivity
0.5 µV/e @ 10000e (Cl. III) OIML R76,
Resolution
7.5 million counts at 3mV/V, not legal for trade
Input voltage (input signal + dead load)
0 ... max. 36 mV DC, symmetrical to 0
Dead load range
36 mV DC (max. input signal); input/ calibration via software