Commissioning
PR 5220 Instrument Manual
EN-68
Sartorius
Weighingpoint/WP A/
Calibration
Tare timeout
2.5 s
Zeroset range
50.00 d
Zerotrack range
0.25 d
Zerotrack step
0.25 d
Zerotrack time
0.0 s
Overload
9 d
[Tare timeout]
Enter a timeout value between 0.1 and 25 s for a taring/zero set command that cannot be executed (e.g., if
scale mechanically unstable, filter settings faulty, resolution too high, stability condition too narrow).
[Zeroset range]
Determine a ±range around the zero point determined by the dead load during calibration; within this range:
-
the displayed gross weight can be set to zero by pressing the zero-setting key (or by a corresponding
external command), and
-
automatic zero tracking is active.
Available range: 0.00 d to 10000.00d
In 'legal-for-trade' mode a value
≤ 2 % of Max. must be entered. Example: 60 d for 3000e, class III.
[Zerotrack range]
Range within which automatic zero tracking compensates deviations; 0.25 to 10000.00d.
In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Zerotrack step]
Step for automatic tracking; 0.25 to 10d. In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Zerotrack time]
Enter a time interval for automatic zero tracking within 0.0 (tracking switched off) and 25 s.
In 'legal-for-trade' mode a value of 0.25 d has to be entered.
[Overload]
Weighing range above the maximum capacity (Max), without error message. Available range: 0 to 9999900 d.
In 'legal-for-trade' mode max. 9 d=e has to be entered.
Weighingpoint/WP A/
Calibration
Minimum weight
50 d
Range mode
Single range
[Minimum weight]
Minimum weight at which a print command can still be executed. Range is 0 to 9999900 d.
In 'legal-for-trade' mode min. 20 d has to be entered.
[Range mode]
For scale range selection, see Chapter 4.4.13.2.