Fieldbus Interface
Sartorius
EN-141
7.4
Fieldbus Register
Register 0: I/O Status Bits for Reading
Dynamic status, only reading is permitted
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Input 3
Input 2
Input 1
Byte 1
Output 3 Output 2 Output 1
Byte 2
Limit 3
Limit 2
Limit 1
Byte 3
Register 1: Scale Status
Dynamic status, only reading is permitted
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
DIM
STND
INZSR
CZERO
BELOW0
OVL
>Max
ADCERR
Byte 1
E6
E1
E3
E7
Byte 2
PowerFail
ActionActive
CmdError
Byte 3
TareActive
CalActive
TestActive
Byte 0 corresponds to byte 7 in the output area; for weight error, see also the table in Chapter 13.1.
ADCERR
Error of analog conversion/ load cell circuit (OR function of the E1, E3, E7 bits)
>Max
Gross value is higher than Max (maximum capacity); the full scale deflection was exceeded.
OVL
Scale overload; Gross >Max + overload,
Error 2
BELOW0
Gross weight is negative (<0-¼ d)
CZERO
Center of zero; weight within 0
±
¼ d
INZSR
The gross weight is within the zero setting range
STND
The scale is stable
DIM
The gross weight has exceeded the full scale deflection (0-¼ d > weight value >Max+overload),
(OR function of bits BELOW0, OVL).
E7
The input signal is negative (inverse conversion)
,
Error 7
E6
No sense voltage, or Sense voltage is low,
Error 6
E3
The input signal is >36mV (no end of conversion),
Error 3
E1
Arithmetic error (overflow),
Error 1
E9
No Communication with xBPI scale,
Error 9
CmdError
Error during execution (cmdError); e.g. the 'taring' operation is not handled, because the scale
is not stable. The error is stored in LastError (register 4). The bit is reset with the ResetError bit
(register 2).
ActionActive The operation is handled; handling is busy
PowerFail
Power failure; is always set after power on. The PowerFail bit is reset with the ResetPWF bit
(register 2) 'Reset power failure'.
TestActive
The analog test is busy
CalActive
The instrument is / was configured. When this bit is 1, the scale parameters (Expo/Unit/Step)
must be read again. Set after power On and reset after reading the max. capacity.
TareActive
The instrument was tared.