Fieldbus interface, Pr 5220 instrument manual – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 144

Fieldbus Interface
PR 5220 Instrument Manual
EN-144
Sartorius
Register 30, 31: Fixed Values (Read/Write)
Register 30
Fixed value for analog output; value (num) 0... 20000 corresponds to 20 mA
Register 31
Fixed value for fixtare; see also SetFixTare, GetFixTare ( see page 142)
Register 80 ... 89: State-Controlled Action Bits (Write)
For setting bits, see page 138.
Only setting and resetting of single bits is possible.
When changing a bit from 0 to 1, the corresponding action starts. After handling the command, the bit must
be reset. Application: the master writes cyclically.
The bit is set as Write_Value_Select with the specified number (see page 138); the bit is reset at the specified
number +128.
Register 80 SetZero
Set the gross weight to zero
Register 81 SetTare
Execute taring
Register 82 ResetTare
Reset tare
Register 83 SetTest
Start the ADC test
Register 84 ResetTest
Finish the ADC test
Register 85 ResetPwf
Reset the PowerFail bit (Register 1; the bit was set after power on)
Register 86 SetFixTare
Taring with weight in numerical address D31 'Fixtare'
Register 87 GetFixTare
The current gross weight is copied into numerical address D31.
Register 89 ResetError
The CmdError error bit is reset
Register 112 ... 121: Transition-Controlled Action Bits (Write)
For setting bits, see page 138.
As soon as the bit was set, it is reset internally and the operation is handled; this is transition-controlled (for
writing once).
The bit is set as Write_Value_Select with the specified number (see page 138).
Register 112
SetZero
Register 113
SetTare
Register 114
ResetTare
Register 115
SetTest
Register 116
ResetTest
Register 117
ResetPwf
Register 118
SetFixTare
Register 119
GetFixTare
Register 121
ResetError