3 weighingpoint ‚ pendeo load cells, Pr 5220 instrument manual commissioning – Rice Lake PR5220 Ethernet Transmitter User Manual
Page 53

PR 5220 Instrument Manual
Commissioning
Sartorius
EN-53
4.3.8.3
Weighingpoint ‚ Pendeo Load Cells’
- Search
Search load cells
-
Search for a new network and reset the load cell data to the factory
settings.
-
Or replace a single faulty load cell with the existing data.
- View
Serial number and current weigh of connected load cells are displayed.
- Info
Data of selected load cell are displayed.
- Assign
Load cells (serial number) are assigned to the mounting place.
- Calib
Calibration xBPI load cells
- New
Corner correction will be reset
Yes, No
- Number of vessel feet
Enter number of vessel feet
- Local gravity
Enter local gravity (Default: Hamburg 9,81379 m/s
2
)
- Max
Max. capacity: 0.000010 ...<3000>... 9999998
- Sccale interval
<1>, 2, 5, 10, 20, 50
- Dead load
Empty scale
- CAL weight
Calibration weight: 0.000010 ... 9999998
- Corner correction
O.k., when realized
- Modify
For minor modifications/ subsequent dead load setting only, otherwise,
use [New].
siehe ‚New‘
- Param
- Ambient conditions
Very stable, stable, unstable, very unstable
- W
& M
- Standstill time
0.01 s…<0.50 s> ... 2.0 s
(The range depends on the measurement time.)
- Standstill range
0.00 d ... <1.00 d> ... 10.00 d
(The range depends on the measurement time.)
- Tare timeout
0.1 s ... <2.5 s> ... 25 s, timeout when there is no standstill
- Zeroset range
0.00 d ... <50.00 d> ... 10000.00 d
- Zerotrack range
0.00 d ... <0.25 d> ... 10000.00 d
- Zerotrack step
0.00 d ... <0.25 d> ... 10.00 d
- Zerotrack time
<0.0 s> ... 25 s
- Overload
0 d ... <9 d> ... 999900 d
- Min
0 d ... <50 d> ... 999900 d, min. weight
- Range mode
- Range limit 1
0…max. weight, transition from small to medium scale interval; only
when multiple range or multi-interval has been selected.
- Range limit 2
0…max. weight, transition from medium to large scale interval; only
when multiple range or multi-interval has been selected.