PR 5220 Instrument Manual
Commissioning
Sartorius
EN-67
4.4.13
Parameter Input
Open the menu via
-[Weighingpoint]-[Calib]-[Param].
Weighingpoint/WP A/
Calibration
Measuretime
160 ms
Digital filter
bessel
External supply
8 - 12 VDC
Fcut
3.00 Hz This line is shown only if the filter is on.
Test mode
absolute
W
& M
none
Standstill time
0.50 s
Standstill range
1.00 d
[Measuretime] 
Select the measuring time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms. 
In 'legal-for-trade' mode select 
≤1 s.
[Digital filter]
The digital filter can be switched on only with the measuring time set to
≤160 ms.
Select the filter characteristic [off], [bessel], [aperiod.], [butterw.], [tscheby.].
[Fcut] 
Enter the cut-off frequency for the filter (0.1…80 Hz); the setting is dependent on the measuring time. 
[External supply] 
With external load cell supply (e.g. 7.5 V DC via 1626/60 with MX8 = closed), [below 8 VDC] can be selected to 
adapt the sense voltage monitoring to the lower supply voltage. 
[Test mode] 
With [Absolute], the test value is calculated when the test is called. With [relative], the deviation from the 
initially stored test value is displayed; see Chapter 4.4.11. 
[W
& M]
Setting for 'legal-for-trade' mode; select [none], [OIML], [NTEP] (for USA) or [NSC] (for Australia), 
see Chapter 4.4.13.1. 
[Standstill time] 
Settings [Standstill range] and [Standstill time] are required for determining the mechanical stability of the 
scale. Input in seconds; permissible range 0.01 bis 2 sec. (max. measuring time x 32). If 0 is set, stability is not 
checked. The stability time must not be less than the measuring time and not greater than 32 times the 
measuring time. 
[Standstill range] 
The scale is stable as long as any changes in the weight value are within this range; permissible range: 
0.01d to 10.00d. In 'legal-for-trade' mode select 
≤1 d.