ITC AVOXimeter 1000E Operator’s Manual
71
Guidance and Manufacturer’s Declaration – Electromagnetic
Emissions
The ITC AVOXimeter 1000E is intended for use in the electromagnetic environment specified
below. The customer or the user of the ITC AVOXimeter 1000E should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The ITC AVOXimeter 1000E uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class B
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
The ITC AVOXimeter 1000E is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.