Requiring entry of a patient id – ITC Avoximeter 1000E Whole Blood Oximeter User Manual
Page 28
ITC AVOXimeter 1000E Operator’s Manual
22
Requiring Entry of a Patient ID
The instrument can be configured so that a Patient ID must be entered before a test can be run.
Important:
The User ID for the “QA User” must be entered in order to enable
the requirement to enter a Patient ID. See page 21 for details.
1.
Display the “User and Patient ID” menu (a submenu of
, “Data Management,”
see page 8).
2.
Press
followed by
to display the current Patient ID setting:
3.
Press
followed by
. A screen is displayed for specifying the minimum
and maximum number of digits:
4.
If the minimum and/or maximum number of digits are to be changed, press
and enter a new value (followed by
) for the minimum and maximum
number of digits.
Note:
If the minimum and/or maximum number of digits is changed, the
instrument will continue to prompt that up to twelve digits is acceptable
when entering a Patient ID. Operators should be instructed to ignore
this prompt.
5.
Respond
to the confirmation prompt. The “User and Patient ID” menu is again
displayed.
6.
Press
to return to the previous menu, if desired.