ITC AVOXimeter 1000E Operator’s Manual
6
2 Description
The ITC AVOXimeter 1000E is a Point of Care device for use at the bedside. It contains a test
chamber which performs all operations to measure the oxyhemoglobin saturation (%Hb0
2
), total
hemoglobin (THb), and oxygen content (O
2
) of a whole blood sample after the operator inserts a
test cuvette containing the sample into the test chamber.
Each ITC AVOXimeter 1000E is calibrated at the factory. The ITC AVOXimeter 1000E can be
operated either from its internal batteries or from the AC adapter. The batteries are charged when
the AC Adapter is connected.
Figure 1. ITC AVOXimeter 1000E Oximeter
Front Panel
The front panel (Figure 2) contains the test chamber, a keypad with the
key, action and
menu keys, number keys, and a display panel. Operator instructions are shown on the display
panel, and the operator enters commands and information using the keypad.
When the test is completed, the results are shown on the display panel and stored in system
memory.
The display panel is illuminated to enhance visibility in low light conditions. The illumination can
be adjusted (or turned off) to conserve power during battery operation.
Test Cuvette
Finger Grip