Running liquid controls – ITC Avoximeter 1000E Whole Blood Oximeter User Manual
Page 41
ITC AVOXimeter 1000E Operator’s Manual
35
Running Liquid Controls
ITC recommends either of the following liquid controls.
●
RNA CO-Oximeter controls, available from RNA Medical, Devens, MA., phone 978 772-
9070 or toll-free 800-533-6162.
●
Multi-4™ CO-Oximeter controls, available from Instrumentation Laboratory, Inc.,
Lexington, MA., phone toll-free 800-955-9525.
1.
Fill a test cuvette with a liquid control (see page 28).
2.
Run a test (see page 29).
3.
Record the THb and %HbO
2
results, or print the results of the test (see page 30).
4.
Verify that the results are within the expected range of values for the liquid control.