ITC Avoximeter 1000E Whole Blood Oximeter User Manual
Page 63
ITC AVOXimeter 1000E Operator’s Manual
57
Symptom
Possible Cause
Action Required
THb and/or %HbO
2
readings for yellow or
orange filter are out of
range.
1. The optical detector may be
contaminated by blood or
foreign material.
1. Clean the optical detector (see
2. The instrument may require
calibration.
2. Re-calibrate the instrument
(see page 39).
The following ERROR:
message is displayed:
“Uninitialized Vector
Service Required“
1. Microprocessor error.
1. Press the
and
keys
simultaneously to turn the
instrument off, then turn the
instrument back on.
2. If the problem persists, contact
ITC Technical Support (see
page ii).