Connections, Automatic standby and shutdown, Instrument lockouts – ITC Avoximeter 1000E Whole Blood Oximeter User Manual
Page 16
ITC AVOXimeter 1000E Operator’s Manual
10
Connections
Connections to the AC Adapter and an optional printer (or a computer) are made at the rear of
the instrument (Figure 4).
Use only the AC Adapter provided with the instrument.
Figure 4. Power and Printer Connections
Automatic Standby and Shutdown
The ITC AVOXimeter 1000E enters a low-power standby mode after the instrument has been idle
for a specified period of time (the standby delay). The instrument is factory preset for a standby
delay time of 60 minutes, but a time of 10 to 180 minutes can be specified (see page 17). To
resume normal operation when the instrument is in standby, press and hold down any key for one
second.
Note:
The ITC AVOXimeter 1000E also enters standby if the battery charge becomes
critically low.
The ITC AVOXimeter 1000E shuts down after it has been in standby for 4 hours.
Instrument Lockouts
The instrument can be configured to allow use only by authorized operators and/or to allow use
only if Optical Quality Control (OQC) has been performed.
Printer or Computer
AC Adapter
Serial No.
Label