Instrument specifications, Reportable range, Accuracy – ITC Avoximeter 1000E Whole Blood Oximeter User Manual
Page 17: Precision, Interference

ITC AVOXimeter 1000E Operator’s Manual
11
Instrument Specifications
Size
20.3 cm (8.0 in) x 25.4 cm (10.0 in) x 9.5 cm (3.8 in)
Weight
1.8 kg (4 lbs)
Operating Temperature
Room temperature (15°C to 30°C, 59°F to 86°F)
Battery Type
Nickel Cadmium (NiCad)
Operating Time On Battery
Approximately 8 hours (constant run at medium brightness).
Tests may also be run while the ITC AVOXimeter 1000E is
plugged into the AC/DC power module.
Anticipated Battery Life
Approximately 500 charge / discharge cycles
Power Supply/Chargers
Input:
100 / 240 VAC, 50 / 60 Hz
Output: 12 VDC, 830 mA
Serial Data Port
RS232C
Sample Type
Whole blood
Sample Volume
50 µL
Analysis Time
7 to 10 seconds per sample
Analysis Wavelengths
5
Reportable Range
%HbO
2
0 to 100%
THb
4 to 25 g/dL
[O
2
Content]
0 to 35 mL O
2
/dL
Accuracy
%HbO
2
±1 %HbO
2
THb (>10 g/dL)
±0.45 g/dL
THb (<10 g/dL)
±0.35 g/dL
Precision
%HbO
2
0.5 %HbO
2
THb
0.3 g/dL
Interference
Bilirubin
None
Hemolysis
None
Carboxyhemoglobin
None
Methemoglobin
< 1% %HbO
2
, < 0.2 g/dL THb
(THb = 16 g/dL, MetHb < 10%, 7.1 < pH < 7.8)
Fetal Hemoglobin
< 1% %HbO
2
, < 0.45 g/dL THb
(THb = 16 g/dL, HbF < 100%)