ITC AVOXimeter 1000E Operator’s Manual
37
Tagging a Liquid Control Test with a Lot Number
The lot number of the control being tested can be included in the results for the test.
Note:
Lot numbers of liquid controls must be entered before they can be tagged to a liquid
QC test (see page 36).
1.
Insert a cuvette filled with liquid quality control material (see page 35). A prompt to
confirm liquid QC is displayed:
2.
Press
to confirm. A prompt to confirm that the level and lot numbers are being
entered is displayed:
3.
Press
to confirm. A menu for selection of a control level is displayed:
4.
Select the level that is being run and press
. A menu for selection of a lot
number is displayed:
5.
Select the previously entered lot number that is being run and press
.
6.
The sample results will be displayed on the screen, and the lot number and LQC
level will be added to the record for that sample.