Bio-Rad Ready Gel Precast Gels for 2-D Electrophoresis User Manual
Page 39

34
Section 12
Troubleshooting
Improper storage of Ready Gel precast gels can produce numerous artifacts. Gels should be stored flat at
4ºC. Avoid freezing or prolonged storage above 4ºC. If you suspect your gels have been stored improperly,
DO NOT USE THEM.
Problem
Possible Cause
Solution
Samples do not migrate into gel
Tape at the bottom of the cassette
Remove tape
not removed
Insufficient buffer in integral buffer chamber
Fill buffer chamber with 125 ml
running buffer
Insufficient lower electrode buffer
Fill both halves of the lower buffer tank
with sufficient running buffer
Electrical disconnection
Check electrodes and connections
Bands “smile” across gel, band pattern
Excess heating of gel
Check buffer composition
curves upward at both sides of the gel
Completely fill both halves of the lower
buffer tank with sufficient running buffer
Do not exceed recommended running
conditions
Skewed or distorted bands, lateral
Excess salt in samples
Remove salts from sample by dialysis or
band spreading
desalting column prior to sample
preparation
Insufficient sample buffer or wrong formulation
Check buffer composition and dilution
instructions
Problem
Possible Cause
Solution
Vertical streaking
Overloaded samples
Dilute sample