Troubleshooting

Improper storage of Ready Gel precast gels can produce numerous artifacts. Gels should be stored flat at
4ºC. Avoid freezing or prolonged storage above 4ºC. If you suspect your gels have been stored improperly,
DO NOT USE THEM.

Problem

Possible Cause

Solution

Samples do not migrate into gel

Tape at the bottom of the cassette

Remove tape

not removed

Insufficient buffer in integral buffer chamber

Fill buffer chamber with 125 ml
running buffer

Insufficient lower electrode buffer

Fill both halves of the lower buffer tank
with sufficient running buffer

Electrical disconnection

Check electrodes and connections

Bands “smile” across gel, band pattern

Excess heating of gel

Check buffer composition

curves upward at both sides of the gel

Completely fill both halves of the lower
buffer tank with sufficient running buffer

Do not exceed recommended running
conditions

Skewed or distorted bands, lateral

Excess salt in samples

Remove salts from sample by dialysis or

band spreading

desalting column prior to sample
preparation

Insufficient sample buffer or wrong formulation

Check buffer composition and dilution
instructions

Problem

Possible Cause

Solution

Vertical streaking

Overloaded samples

Dilute sample