3.2 Remote control commands
758 KFD Titrino
89
"Configuration",
continuation
├
.RSSet1
Settings RS232, 1
$G 3.2.2.79.
│ ├
.Baud Baud
rate
300,600,1200,2400,4800,
│
│
9600,19200,38400,57600,
│
│
115200
ditto
│ ├
.DataBit
Number of data bits
7, 8 ditto
│ ├
.StopBit
Number of stop bits
1, 2
ditto
│ ├
.Parity Parity
even,
odd,
none
ditto
│ ├
.Handsh Handshake
HWs, SWchar,
│
│
SWline, none
ditto
├
.RSSet2
as for RS1
│
├
.ComVar
Values of common variables
│ ├
.C30 C30
0... ±999999
3.2.2.80.
│ ├
up to C39
0... ±999999
│
├
.DosPrep
Parameters for the preparation of burets
│ ├
.PowerOnPrep
Warning after power on
ON, OFF 3.2.2.81.
│ ├
.Report Preparation
report
ON, OFF 3.2.2.82.
│ ├
.Select
Selection of dosing unit
internal D0, external D1,
│
external
D2
3.2.2.83.
│ ├
.D0
Internal Titrino buret
│
│ ├
.WarnInterval Warning interval for preparation
5...9999, OFF 3.2.2.84.
│
│ ├
.V Volume
0...3.5...99999.99 3.2.2.85.
│
│ ├
.Repeat
Number of cycles
1...2...9 3.2.2.86.
│
│ ├
.DosRate Dosing
rate
0.01...150,
max. 3.2.2.87.
│
│ ├
.FillRate Filling
rate
0.01...150,
max. ditto
│ ├
.D1
Buret D1
│
│ ├
.WarnInterval Warning interval for preparation
5...9999, OFF 3.2.2.84.
│
│ ├
.Select
Type of dosing unit
Dosimat, Dosino
3.2.2.88.
│
│ ├
.Dosimat
Parameters for Dosimats
│
│
│ ├
.V Volume
0...3.5...99999.99 3.2.2.85.
│
│
│ ├
.Repeat
Number of cycles
1...2...9 3.2.2.86.
│
│
│ ├
.DosRate Dosing
rate
0.01...150,
max. 3.2.2.87.
│
│
│ ├
.FillRate Filling
rate
0.01...150,
max. ditto
│
│ ├
.Dosino
Parameters for Dosinos
│
│
│ ├
.Outlet Expelling
outlet
tip, flask
3.2.2.89.
│
│
│ ├
.DosTubing Size of dosing tubes
│
│
│
│ ├
.Length Length
1...40.0...999.9 3.2.2.90.
│
│
│
│ ├
.Diam Diameter
0.1...2.0...9.9 ditto
│
│
│ ├
.AspirTubing Size of aspirating tubes
│
│
│
│ ├
.Length Length
1...25.0...999.9 ditto
│
│
│
│ ├
.Diam Diameter
0.1...2.0...9.9 ditto
│
│
│ ├
.DosRate Dosing
rate
0.01...150,
max. 3.2.2.91.
│
│
│ ├
.FillRate Filling
rate
0.01...150,
max. ditto
│ ├
.D2
Buret D2 as for D1