3.2 Remote control commands
758 KFD Titrino
122
3.2.2.115. Info.ActualInfo.Oven.HeatTime read
only
Info.ActualInfo.Oven.SampleTemp read
only
Info.ActualInfo.Oven.LowTemp read
only
Info.ActualInfo.Oven.HighTemp read
only
Info.ActualInfo.Oven.GasFlow read
only
Info.ActualInfo.Oven.UnitFlow read
only
$Q sends the current values from a connected KF Oven. If no Oven is con-
nected, the values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.
.GasFlow: Average gas flow during sample heating time.
.UnitFlow: Unit of gas flow.
3.2.2.116. Info.ActualInfo.Display.L1
up to 32 ASCII characters
Info.ActualInfo.Display.L8
up to 32 ASCII characters
Info.ActualInfo.Display.DelAll $G
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 3.2.2.132.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.132.
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.76) to refresh the
display.
$Q sends the contents of the corresponding display line.
3.2.2.117. Info.ActualInfo.Comport.Number read
only
$Q sends the comport number of the Titrino where the PC is connected.
3.2.2.118. Info.Assembly.CycleTime read
only
Info.Assembly.ExV read
only
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.
3.2.2.119. Assembly.Bur.Select internal
D0, external D1, external D2
Selection of buret for assembly functions.
3.2.2.120. Assembly.Bur.Empty
$G, $S, $H, $C
Starts the function "empty". Only possible with the burette type "Dosino". The
parameters are under the function &Config.DosPrep, see 3.2.2.89 up to
3.2.2.91.