Metrohm 758 KFD Titrino User Manual
Page 79
3.1 General rules
758 KFD Titrino
75
.Req .Id1:
Instrument in the TIP mode, requesting Id1 after start.
.Id2:
Instrument in the TIP mode, requesting Id2 after start.
.Id3:
Instrument in the TIP mode, requesting Id3 after start.
.Smpl:
Instrument in the TIP mode, requesting sample size after start.
.Unit:
Instrument in the TIP mode, requesting unit of sample size after start.
.Pause:
Instrument in the TIP mode, in pause.
.Info:
Instrument in the TIP mode, in info.
.Mode...:
Instrument in the TIP mode, working off a submethod. The detailed
status messages of the submethod appear, see above.
Status conditions of the global $H:
The status message of the action which has been held appears.
If the process is held because a monitored limit has been violated, its status
message is $H.Mode.XXX.Titr.
Status conditions of the global $C:
The status conditions of the global $C are identical with the ones of the global
status $G. They appear when the process has been restarted actively from the
status "Hold" ($C, key
error).
Status conditions of the global $R:
$R .Mode.XXXX.QuickMeas:
Quick manual measurement from the initial status in mode XXXX.
$R .Mode.KFT .Inac:
Instrument in the KFT mode, inactive.
.Cond.Ok:
Instrument in the KFT mode, conditioning, endpoint reached.
.Cond.Prog:
Instrument in the KFT mode, conditioning, endpoint not reached.
$R .Mode.SET...
As KFT.
$R .Mode.MEAS .Inac:
Instrument in the MEAS mode, inactive.
$R .Mode.CAL .Inac:
Instrument in the CAL mode, inactive.
$R .Assembly.Bur.ModeDis:
Buret in the DIS mode, inactive.
$R .TIP.Inac:
Instrument in TIP, inactive.
Status conditions of the global $S:
$S .Mode.XXXX.QuickMeas:
Quick manual measurement from the initial status in mode XXXX.
The instrument gives the status from which it has been stopped. The detailed
status information is therefore identical the information for the global status
$G.
Violation of monitored limits with action "end" give the status message
$S.Mode.XXX.Inac;EYYY.