Metrohm 758 KFD Titrino User Manual
Page 111
3.2 Remote control commands
758 KFD Titrino
107
3.2.2.44. Mode.Parameter.Calibration.ActPulse first,
all, OFF
Parameters for CAL: Output of a pulse on the I/O line "Activate", see page
174.
3.2.2.45. Mode.Parameter.Sequence.X.Select
method, pause, L4 output,
L6 output, info, prep, stirrer, OFF
Parameters for TIP: Selection of an element for step X (X = 1...30). For the
parameters of the elements see 3.2.2.46.
3.2.2.46. Mode.Parameter.Sequence.X.Method Method
name
Mode.Parameter.Sequence.X.Pause
0...999999, INF
Mode.Parameter.Sequence.X.L4Output
active, inactive,
pulse, OFF
Mode.Parameter.Sequence.X.L6Output
as for L4
Mode.Parameter.Sequence.X.Info
up to 16 ASCII characters
Mode.Parameter.Sequence.X.Prep
internal D0, external D1,
external
D2
Mode.Parameter.Sequence.X.Stirrer
ON, OFF
Parameters for TIP: Parameters of the elements of TIP.
.Method: Method name of a method available in the user memory or on the
card. Up to 8 ASCII characters.
.Pause:
Pause time in s. INF means infinite. Continue the sequence with
&m $G.
.L4 Output: Warning: A pulse triggered by the limit value monitoring at L4 (pin
3) in a submethod sets an output set to active in TIP to inactive.
.L6 Output: Warning: An activate pulse at L6 output (pin 1) in a submethod
sets an output set to active in TIP to inactive.
.Info:
Entry of a message which is written into the display. The se-
quence remains in the display with the corresponding message.
Continue with &m $G.
.Prep:
Preparation of titrating burette.
.Stirrer:
Switching stirrer on/off.
3.2.2.47. Mode.Parameter.Presel.MeasMode
pH, U, Ipol, Upol, OFF
Mode.Parameter.Presel.MeasInput
1, 2, diff.
Mode.Parameter.Presel.Ipol 0...1...±127
Mode.Parameter.Presel.Upol 0...400...±1270
Mode.Parameter.Presel.PolElectrTest ON, OFF
Mode.Parameter.Presel.Temp –170....25.0...500.0
Parameters for TIP: Selection of the measured quantity for manual measure-
ments in the inactive state, see 3.2.2.2. Selection of the measuring input
(MeasInput) applies to measured quantities pH and U. "diff." means differen-
tial amplifier, see page 167. With Ipol the requests for the polarization current
in µA (Ipol) and .PolElectrTest apply. With Upol the request for the polarization
voltage in mV (Upol) applies. Entry in steps of 10 mV. .PolElectrTest also ap-
plies. If the test for polarized electrodes is switched on (ON), it will be per-
formed on the change from the inactive state to an active state. The tempera-
ture applies to pH measurements.