beautypg.com

Metrohm 758 KFD Titrino User Manual

Page 111

background image

3.2 Remote control commands

758 KFD Titrino

107

3.2.2.44. Mode.Parameter.Calibration.ActPulse first,

all, OFF

Parameters for CAL: Output of a pulse on the I/O line "Activate", see page

174.

3.2.2.45. Mode.Parameter.Sequence.X.Select

method, pause, L4 output,

L6 output, info, prep, stirrer, OFF

Parameters for TIP: Selection of an element for step X (X = 1...30). For the

parameters of the elements see 3.2.2.46.

3.2.2.46. Mode.Parameter.Sequence.X.Method Method

name

Mode.Parameter.Sequence.X.Pause

0...999999, INF

Mode.Parameter.Sequence.X.L4Output

active, inactive,

pulse, OFF

Mode.Parameter.Sequence.X.L6Output

as for L4

Mode.Parameter.Sequence.X.Info

up to 16 ASCII characters

Mode.Parameter.Sequence.X.Prep

internal D0, external D1,

external

D2

Mode.Parameter.Sequence.X.Stirrer

ON, OFF

Parameters for TIP: Parameters of the elements of TIP.

.Method: Method name of a method available in the user memory or on the

card. Up to 8 ASCII characters.

.Pause:

Pause time in s. INF means infinite. Continue the sequence with

&m $G.

.L4 Output: Warning: A pulse triggered by the limit value monitoring at L4 (pin

3) in a submethod sets an output set to active in TIP to inactive.

.L6 Output: Warning: An activate pulse at L6 output (pin 1) in a submethod

sets an output set to active in TIP to inactive.

.Info:

Entry of a message which is written into the display. The se-

quence remains in the display with the corresponding message.

Continue with &m $G.

.Prep:

Preparation of titrating burette.

.Stirrer:

Switching stirrer on/off.

3.2.2.47. Mode.Parameter.Presel.MeasMode

pH, U, Ipol, Upol, OFF

Mode.Parameter.Presel.MeasInput

1, 2, diff.

Mode.Parameter.Presel.Ipol 0...1...±127

Mode.Parameter.Presel.Upol 0...400...±1270

Mode.Parameter.Presel.PolElectrTest ON, OFF

Mode.Parameter.Presel.Temp –170....25.0...500.0

Parameters for TIP: Selection of the measured quantity for manual measure-

ments in the inactive state, see 3.2.2.2. Selection of the measuring input

(MeasInput) applies to measured quantities pH and U. "diff." means differen-

tial amplifier, see page 167. With Ipol the requests for the polarization current
in µA (Ipol) and .PolElectrTest apply. With Upol the request for the polarization

voltage in mV (Upol) applies. Entry in steps of 10 mV. .PolElectrTest also ap-

plies. If the test for polarized electrodes is switched on (ON), it will be per-

formed on the change from the inactive state to an active state. The tempera-

ture applies to pH measurements.