Metrohm 758 KFD Titrino User Manual
Page 88
![background image](/manuals/586878/88/background.png)
3.2 Remote control commands
758 KFD Titrino
84
*Parameter
Tree part "Parameters for MEAS"
├
.
Measuring
Measuring parameters
│ ├
.SignalDrift
Drift for meas.value acquisition
depends on meas.quant. 3.2.2.34.
│ ├
.UnitSigDrift
Unit of measured value drift
read only
ditto
│ ├
.EquTime Equilibrium
time
0...9999,
OFF ditto
│ ├
.MeasInput Measuring
input
1, 2, diff.
3.2.2.35.
│ ├
.Ipol
Polarization
current
0...1...±127 ditto
│ ├
.Upol
Polarization voltage
0...400...±1270 ditto
│ ├
.PolElectrTest
Test for polarized electrodes
ON, OFF ditto
│ ├
.Temp
Titration temperature
–170.0...25.0...500.0 3.2.2.36.
│ ├
.TDelta
Time interv.for meas.acquisition
1...2...999999 3.2.2.37.
│ │
├
.
Statistics
Statistics
│ ├
.Status
Status of statistics calculation
ON, OFF 3.2.2.26.
│ ├
.MeanN
No. of individual determinations
2...20 ditto
│ ├
.ResTab Result
table
│ │ ├
.Select
original,delete n,delete all
ditto
│ │ ├
.DelN
Deletion of individual results
1...20 ditto
│ │
├
.
Presel
Preselections
│ ├
.IReq
Request of Id's after start
id1, id1&2, all, OFF 3.2.2.28.
│ ├
.SReq
Request of sample size after start
value, unit, all, OFF ditto
│ ├
.LimSmplSize
Limits for sample size
3.2.2.29
│ │ ├
.Status
Status of limit control
ON, OFF ditto
│ │ ├
.LoLim Lower
limit
0.0...999 999
ditto
│ │ ├
.UpLim Upper
limit
0.0...999 999 ditto
│ ├
.ActPulse
Output of a pulse
ON, OFF 3.2.2.30.
*Parameter
Tree part "Parameters for CAL"
├
.
Calibration
Calibration parameters
│ ├
.MeasInput Measuring
input
1, 2, diff.
3.2.2.38.
│ ├
.CalTemp Calibration
temperature –20.0...25.0...120.0 3.2.2.39.
│ ├
.Buffer
│ │ ├
.1
│ │ │ ├
.Value
pH value of buffer 1
0...7.00...±20.00 3.2.2.40.
│ │ ├
.2
│ │ │ ├
.Value
pH value of buffer 2
0...4.00...±20.00, OFF
ditto
│ │ ├
:
up to 9 buffers
│ ├
.SignalDrift
Drift for meas.value acquisition
0.5...2...999, OFF
3.2.2.41.
│ ├
.EquTime Equilibrium
time
0...100...9999, OFF
ditto
│ ├
.ElectrodeId
Electrode identification
8 ASCII char.
3.2.2.42.
│ ├
.SmplChanger
Calibration on a Titrino
ON, OFF 3.2.2.43.
│ ├
.ActPulse
Output of a pulse
first, all, OFF 3.2.2.44.
│ │
├
.
Statistics
Statistics
│ ├
.Status
Status of statistics calculation
ON, OFF 3.2.2.26.
│ ├
.MeanN
No.of individual determinations
2...20 ditto
│ ├
.ResTab Result
table
│ │ ├
.Select
original,delete n,delete all
ditto
│ │ ├
.DelN
Deletion of individual results
1...20 ditto