Metrohm 758 KFD Titrino User Manual
Page 105
3.2 Remote control commands
758 KFD Titrino
101
3.2.2.11. Mode.CFmla
Mode.CFmla.1.Value
0...±999999
Mode.CFmla.2.Value
etc., up to .19
Calculation constants specific to a method. Stored in the method memory of
the Titrino. Operands specific to the sample (3.2.2.93 and 3.2.2.106) and val-
ues of common variables (3.2.2.80) on the other hand are not stored with the
methods.
3.2.2.12. Mode.Parameter.SET1.EP pH:
0...±20.00, OFF
U,
Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Mode.Parameter.SET1.UnitEp read
only
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.13. Mode.Parameter.SET1.Dyn pH:
0.01...20.00, OFF
U,
Ipol:
1...2000, OFF
Upol:
0.1...200.0, OFF
Mode.Parameter.SET1.UnitDyn read
only
Mode.Parameter.SET1.MaxRate 0.01...10...150, max.
Mode.Parameter.SET1.MinRate 0.01...25.0...999.9
Parameters for SET: Control parameters, see page 33.
.Dyn:
Dynamics, control range in pH, mV (with U and Ipol) or µA (with
Upol). The corresponding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with the Exchange Unit in current use.
.MinRate: Minimum titration rate in µL/min.
3.2.2.14. Mode.Parameter.SET1.Stop.Type drift, time
Mode.Parameter.SET1.Stop.Drift 1...20...999
Mode.Parameter.SET1.Stop.Time 0...10...999, inf.
Mode.Parameter.SET1.Stop.StopT 0...99999, OFF
Parameters for SET, KFT: Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in µL/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected.
"inf." means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".
3.2.2.15. Mode.Parameter.TitrPara.Direction +,
–, auto
Parameters for SET, KFT: Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment. If 2 EP's have been set in a SET titration, the titration direction is given
by the two EP's. The entry of the titration direction is then invalid.