Metrohm 758 KFD Titrino User Manual
Page 108
3.2 Remote control commands
758 KFD Titrino
104
3.2.2.27. Mode.Parameter.Presel.Cond ON, OFF
Mode.Parameter.Presel.DriftDisp ON, OFF
Mode.Parameter.Presel.DCor.Type auto,
man., OFF
Mode.Parameter.Presel.DCor.Value 0.0...99.9
Parameters for SET, KFT:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
3.2.2.28. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Parameters for SET, KFT, MEAS: Automatic inquiry after the start of the de-
termination. From such an inquiry, the determination continues if the re-
quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.93)
or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.29. Mode.Parameter.Presel.LimSmplSize.Status ON,
OFF
Mode.Parameter.Presel.LimSmplSize.LoLim
0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999
Limit control for the sample size.
3.2.2.30. Mode.Parameter.Presel.ActPuls ON, OFF
for SET, KFT: first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 174.
3.2.2.31. Mode.Parameter.CtrlPara.EP pH:
0...±20.00, OFF
U,
Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Mode.Parameter.CtrlPara.UnitEp read
only
Parameters for KFT: Setting of the EP, resp. Control point as a pH value, in
mV (with U and Ipol) or µA (with Upol). The corresponding unit can be read
with .UnitEP.
With KFT the standard value for Ipol is 250 mV, for Upol 25 µA.