3.2 Remote control commands

758 KFD Titrino

104

3.2.2.27. Mode.Parameter.Presel.Cond ON, OFF

Mode.Parameter.Presel.DriftDisp ON, OFF

Mode.Parameter.Presel.DCor.Type auto,

man., OFF

Mode.Parameter.Presel.DCor.Value 0.0...99.9

Parameters for SET, KFT:

.Cond: Conditioning ON/OFF

.DriftDisp: Drift display during conditioning ON/OFF.

.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of

the drift value at start.

.DCor.Value: Drift value for the manual drift correction.

3.2.2.28. Mode.Parameter.Presel.IReq

id1, id1&2, all, OFF

Mode.Parameter.Presel.SReq

value, unit, all, OFF

Parameters for SET, KFT, MEAS: Automatic inquiry after the start of the de-

termination. From such an inquiry, the determination continues if the re-

quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.93)

or with &M $G, see 3.2.2.1.

$H is not possible in requests.

3.2.2.29. Mode.Parameter.Presel.LimSmplSize.Status ON,

OFF

Mode.Parameter.Presel.LimSmplSize.LoLim

0.0...999 999

Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999

Limit control for the sample size.

3.2.2.30. Mode.Parameter.Presel.ActPuls ON, OFF

for SET, KFT: first, all, cond., OFF

Output of a pulse on the I/O line "Activate", see page 174.


3.2.2.31. Mode.Parameter.CtrlPara.EP pH:

0...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0...±200.0, OFF

Mode.Parameter.CtrlPara.UnitEp read

only

Parameters for KFT: Setting of the EP, resp. Control point as a pH value, in
mV (with U and Ipol) or µA (with Upol). The corresponding unit can be read

with .UnitEP.
With KFT the standard value for Ipol is 250 mV, for Upol 25 µA.