3.1 General rules
758 KFD Titrino
78
E135
Check temp.sensor in MEAS T or with activated temperature monitoring.
Exit: Correct error or &m $S.
E136
Same buffer in CAL. Measured value of the second buffer differs less than
6 mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
E137
XXX Bytes are missing so that the method or the silo line could not be stored
or not enough RAM for running TIP.
Exit: Send new command.
E155
No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
E157
No sequence defined in TIP.
Exit. The error message disappears on next start.
E158
A second TIP has been called up in TIP.
Exit: The error message disappears on next start.
E160
No new temporary variable.
Exit: The error message disappears on next start.
E161
Measurement range of the secondary measured value (temperature)
exceeded. The primary measured value (pH, U, Ipol, Upol) can also be
unstable.
Exit: Rectify error or &m $S.
E166
Save lines is "OFF" although a submethod of TIP includes an assignment to
C24 or C25.
Exit: The error message disappears on next start. Attention: The data of this
sample will not be stored.
E172
In TIP a QuickMeas was started, without defining a measuring quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
E173
The warning interval of the internal buret D0 called.
Exit: Execute prep &a.b.p $G or start next method.
E174
The warning interval of the external buret D1 called.
Exit: Execute prep &a.b.p $G or start next method.
E175
The warning interval of the external buret D2 called.
Exit: Execute prep &a.b.p $G or start next method.
E176
The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was
interrupted manually.
Exit. The error message disappears on next start.
E177
Accessing to the memory card, the card was not (properly) inserted.
Exit: The error message disappears on next start.
E178
The date of changing the battery of the card is expired.
Exit: The error message disappears on next start.
E180
Memory card write-protected.
Exit: Send new command.
E181
Memory card not formatted.
Exit: Send new command.
E182
Memory card not accessible.
Exit: Send new command.
E183
A directory with the same name exists already on the memory card.
Exit: Send new command.
E196
Result is out of limits.
Exit: The error message disappears on next start or on recalculation.