3.2 Remote control commands
758 KFD Titrino
102
3.2.2.16. Mode.Parameter.TitrPara.XPause
0...999999
Parameter for SET, KFT: Pause time in s. Runs before dosing the start vol-
ume.
3.2.2.17. Mode.Parameter.TitrPara.StartV.Type abs.,
rel.,
OFF
Mode.Parameter.TitrPara.StartV.V 0...999.99
Mode.Parameter.TitrPara.StartV.Factor
0...±999999
Mode.Parameter.TitrPara.StartV.Rate 0.01...150, max.
Parameters for SET, KFT: Start volume.
If an absolute start volume (abs.) has been selected, the volume in mL is
valid.
A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size ∗ factor
The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.
3.2.2.18. Mode.Parameter.TitrPara.Pause
0...999999
Parameters for SET, KFT: Pause time in s. Is waited off after the dispensing of
the start volume.
3.2.2.19. Mode.Parameter.TitrPara.ExtrT
0...999999
Parameter for SET, KFT: Extraction time in s.
3.2.2.20. Mode.Parameter.TitrPara.DosUnit internal
D0, external D1/D2
Mode.Parameter.DosPara.DosUnit internal
D0, external D1/D2
.TitrPara.DosUnit: Parameter for SET, KFT. Selection of dosing unit.
3.2.2.21. Mode.Parameter.TitrPara.MeasInput
1, 2, diff.
Mode.Parameter.TitrPara.Ipol –127...1...+127
Mode.Parameter.TitrPara.Upol –1270...400...+1270
Mode.Parameter.TitrPara.PolElectrTest ON, OFF
Parameters for SET, KFT:
Selection of the measuring input; valid with measured quantities pH and U.
"diff." means differential amplifier, see page 167.
With Ipol, the inquiries for the polarization current in µA (Ipol) and
.PolElectrTest are valid. Standard for Ipol with KFT is 50 µA.
With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry in
steps of 10 mV.
Besides .PolElectrTest is valid.
If the test for polarized electrodes is switched on, it is performed on change-
over from the inactive state to an active state (titration or conditioning).